Adequate visualization of the retina is critical to the success of vitreoretinal surgery. The advent of endoillumination represents a substantial improvement in the way that surgeons are able to visualize the retina because, compared with prior surgical viewing modalities, it decreases the reflections and scattering from ocular structures such as the cornea, lens, intraocular lens implant, and vitreous.

Chandelier lighting systems provide a stationary and diffuse form of endoillumination from an anterior location. 1 Currently, several companies manufacture chandelier lighting systems, including Insight Instruments with the Tornambe Torpedo,2 Alcon, Synergetics, and DORC. Typically, these products are available in 23- and 25-gauge formats.3 More recently, 27- and 29-gauge models have become available.4,5 Some models of chandelier light probes may be placed into traditional trocar systems (Figures 1 and 2), while others require a separate tool to create the sclerotomy incision. The chandelier systems currently available are either singleor dual-fiber systems, and some provide simultaneous infusion.

Advantages

Chandeliers offer several advantages over the use of a separate endoilluminator probe. The principal advantage is that chandeliers enable bimanual surgery because one hand of the surgeon is not required to hold the endoilluminator. The surgeon is able to use the vitreous cutter or scissors in 1 hand and forceps in the other hand. This affords the surgeon more versatility during surgery and is preferred by some surgeons in complex cases such as proliferative vitreoretinopathy, giant retinal tears, intraocular foreign body, pars plana lensectomy for dropped nucleus, and tractional retinal detachments.

The diffuse illumination provided by chandeliers can enhance visualization and facilitate the recording of surgical videos by providing constant, consistent illumination during the case. In addition, the chandelier lighting system enables the surgeon to complete many cases without the need of an assistant. One may also consider using a chandelier when performing endoscopic surgery.

Phototoxicity

With any illumination system applied to the retina, including with the use of a chandelier lighting system, there is a potential risk of phototoxicity.6 Damage to the retina by illumination has been documented for decades in both anterior segment and vitreoretinal procedures.7-9 Light-induced thermal damage to the retina, however, is dependent on several factors. These include the light source (for example, metal halide, halogen, xenon, or mercury vapor), the use of any wavelength barrier filters, the luminous flux (commonly referred to as brightness, measured in lumens), the distance of the light from the retina, and the duration of exposure.

Chandeliers have large divergence angles, spreading the light source out over a larger surface area and reducing irradiance. Irradiance is the power of the electromagnetic radiation per unit area and is typically given in the unit watts/m2. The decreased irradiance theoretically decreases the risk of phototoxicity to the retina and offers another advantage of a chandelier lighting system in vitreoretinal surgery.6 This theory has also been tested experimentally.10 This reduction occurs because the working distance of the irradiance from the light source to the retina is much larger. Increasing the working distance of the light source increases the safety of the light. It is important to note, however, that chandelier illumination is immobile and does not allow the retina time to recover during periods of inactivity, such as occurs with endoillumination.

Disadvantages

Despite its advantages, there may also be several shortcomings to this illumination system. The insertion of a chandelier light requires at least 1 extra sclerotomy into the eye. The additional sclerotomy entails the typical risks of creating any sclerotomy, including leakage, hypotony, infection, and dislocation of the light during the procedure.

With chandelier illumination, visualization of the retina may be compromised. When using a single-fiber chandelier, there may be “shadowing” by instruments, which may interfere with the surgical field. In certain cases, the chandelier does not illuminate the entire field and, therefore, may require repositioning during the surgery. Chandelier illumination can also eliminate the surgeon's ability to reflect the light from an endoilluminator off the vitrectomy cutter to produce a form of retroillumination. Consequently, the diffuse illumination may reduce the ability to see retinal structures clearly.

Another possible disadvantage of chandelier illumination occurs during fluid-air exchange, when the lighting may produce a substantial amount of glare. This often requires either repositioning or turning off the light. Chandeliers may also create safety issues. There have been cases reported in which the chandelier may become very hot and risk thermal damage to the retina, when the chandelier probes are under air11 or when exposed to uveal tissue or hemorrhage.12

Bimanual Surgery

In certain surgical circumstances, surgeons may prefer bimanual surgery rather than having one hand occupied with an endoilluminator probe. In cases of retinal detachment with proliferative vitreoretinopathy or tractional retinal detachments, it is useful to have 2 hands to aid in peeling fibrotic membranes. One hand may use forceps to peel the membranes, and the other may be able to provide counter-traction with a second instrument, such as a pick or even the vitrector. In cases of an intraocular foreign body, a chandelier light allows 1 hand to grab the foreign body with forceps while the second hand supports the elevation of the mass. In patients with a dropped nucleus, the vitrector or fragmatome may be used to elevate the nucleus and the second instrument can be used to chop the nucleus into smaller pieces to facilitate removal by the vitrector or fragmatome. In patients with a giant retinal tear, chandelier illumination provides diffuse illumination to detect the edges of the pathology or the presence of additional retinal tears.

Summary

The use of a chandelier in vitreoretinal surgery is dictated by surgeon preference. It is clear, however, that there are many potential surgical scenarios in which this form of endoillumination is beneficial.

Matthew T. Witmer, MD, is a second-year vitreoretinal fellow at Weill Cornell Medical College in New York.

R.V. Paul Chan, MD, is an Associate Professor of Ophthalmology, the St. Giles Associate Professor of Pediatric Retina, and Director of the Retina Service at Weill Cornell Medical College. Dr. Chan states that he has no financial interests or relationships to disclose. He can be reached via email at roc9013@med.cornell.edu.

Dean Eliott, MD, is Associate Director of the Retina Service, Massachusetts Eye and Ear Infirmary, Harvard Medical School, and is a Retina Today Editorial Board member. He may be reached by phone: +1 617 573-3736; fax: +1 617 573- 3698; or via email at dean_eliott@meei. harvard.edu.

Ingrid U. Scott, MD, MPH, is a Professor of Ophthalmology and Public Health Sciences, Penn State College of Medicine, Department of Ophthalmology, and is a Retina Today Editorial Board member. She may be reached by phone: +1 717 531 8783; fax: +1 717 531 5475; or via email at iscott@psu.edu.

  1. Zivojnovic R. Silicone Oil in Vitreoretinal Surgery. Dordrecht, Netherlands: Martinus Nijhoff/ Dr W Junk Publishers; 1987.
  2. Tornambe P. Vitrectomy surgery illumination using #26 torpedo mini lights. Video presented at: The Fourth Annual Vitreous Society Film Festival; April 2-4, 2002; San Francisco, CA.
  3. Eckardt C. Twin lights: A new chandelier illumination for bimanual surgery. Retina. 2003;23(6):893-894.
  4. Eckardt C, Eckert T, Eckardt U. 27-gauge Twinlight chandelier illumination system for bimanual transconjunctival vitrectomy. Retina. 2008;28(3):518-519.
  5. Sakaguchi H, Oshima Y, Nishida K, et al. A 29/30-gauge dual-chandelier illumination system for panoramic viewing during microincision vitrectomy surgery. Retina. 2011;31(6):1231-1233.
  6. Charles S. Illumination and phototoxicity issues in vitreoretinal surgery. Retina. 2008;28(1):1-4.
  7. McDonald HR, Irvine AR. Light-induced maculopathy from the operating microscope in extracapsular cataract extraction and intraocular lens implantation. Ophthalmology. 1983;90(8):945-951.
  8. Fuller D, Machemer R, Knighton RW. Retinal damage produced by intraocular fiber optic light. Am J Ophthalmol. 1978;85(4):519-537.
  9. Meyers SM, Bonner RF. Retinal irradiance from vitrectomy endoilluminators. Am J Ophthalmol. 1982;94(1):26-29.
  10. Vander JF, Eagle RC Jr, Brown GC, et al. Retinal tolerance of an implantable light source for use during vitrectomy surgery. Ophthalmic Surg. 1991;22(12):735-739.
  11. Shimada H, Nakashizuka H, Hattori T, et al. Thermal injury caused by chandelier fiber probe. Am J Ophthalmol. 2007;143(1):167-169.
  12. Mathias M, Ernst BJ, Pichi F, et al. Thermal deformation of chandelier endoillumination probes exposed to uveal tissue and blood. Retina. 2012;32(4):773-775.