KEY TAKEAWAYS
- Anatomy (ie, endothelial status, chamber depth, iris quality, and posterior segment conditions) accounts for most of the surgical decision making when choosing the right secondary IOL technique.
- Intraoperative challenges remain significant as secondary IOL techniques rely on ad hoc modifications and off-label use of lenses, leading to breakage and distortion.
- A prosthetic capsular bag (LensOne, Long Bridge Medical) recently received FDA Breakthrough Device Designation.
Secondary IOL implantation has become common in most vitreoretinal surgical practices. Late in-the-bag dislocation is seen more often, driven by an aging pseudophakic population, earlier cataract surgery, longer surgical lifespans, and changes to IOL designs and surgical technique.1 Long-term data suggest a cumulative incidence of secondary IOL implantation of approximately 2% in eyes without pseudoexfoliation, rising to approximately 6% in those with pseudoexfoliation.2 The indications themselves—trauma, complicated cataract surgery, capsular instability—haven’t changed much, but how surgeons approach them has.
A recent review emphasizes that long-term visual outcomes are broadly similar across anterior chamber, iris-fixated, and scleral-fixated IOLs when selection is appropriate.3,4 Thus, the decision is now less about ranking techniques and more about fitting the approach to the eye and the surgical context.
CONSIDERATIONS IN TECHNIQUE SELECTION
Anatomy—including endothelial status, chamber depth, iris quality, and posterior segment conditions—accounts for most of the surgical decision making when choosing the right secondary IOL technique.
In addition, prior ocular surgeries can substantially alter the risk profile.5 A straightforward aphakic eye is not the same as a traumatized, previously vitrectomized eye with glaucoma. As a result, these cases are increasingly managed in combination with anterior segment surgeons. In practice, surgeon familiarity with fixation techniques and available IOL platforms still matters, particularly when intraoperative decisions must be made quickly.
ANTERIOR-CHAMBER IOLS
Modern open-loop anterior-chamber IOLs (ACIOLs) are the only FDA-approved solution for aphakia without capsular support. An ACIOL remains a potential option for an eye with a healthy endothelium, deep chamber, intact angle, and no other considerations that might make the patient a poor candidate. However, many eyes requiring secondary IOL implantation have complex ocular histories or may require future anterior segment intervention, making ACIOLs less suitable in select patients. Although ACIOLs are a common secondary IOL procedure, recent US vitreoretinal practice data suggest an increasing use of scleral-fixated and scleral-sutured techniques.6
ACIOL sizing is a major driver of outcomes. Oversizing can crowd the angle or cause iris tuck, while undersizing leads to movement and endothelial stress. Even with improved lens designs, endothelial reserve still limits long-term tolerance in borderline eyes.7 The limited range of available ACIOL sizes—and variations in OR inventory—may constrain optimal size selection, particularly in unplanned cases.8
IRIS-FIXATED IOLS
Iris fixation includes iris-claw lenses and sutured iris fixation techniques.9 Both avoid scleral dissection and allow posterior chamber positioning. The success of these approaches depends heavily on the quality and stability of the iris tissue; trauma, atrophy, or chronic inflammation can make fixation unreliable. Even in good candidates, pigment dispersion and low-grade inflammation are possible, and patients may report glare or subtle dysphotopsias.10
No iris-claw lenses are approved for the treatment of aphakia in the United States. Outside of the United States, iris-claw options include the Artisan Phakic and Aphakia IOLs (Ophtec).11 The Verisyse Phakic IOL (Johnson & Johnson Vision), marketed in the United States for refractive correction of myopia, is not an FDA-approved aphakic secondary IOL.12
The technical challenges of iris suturing, the need for suitable iris tissue, and the risks of inflammation or pupil-related symptoms have led to limited use in the United States.
SCLERAL-FIXATED IOLS
Sutured scleral fixation (SSF) and sutureless intrascleral haptic fixation (ISHF) share an endpoint but not a mechanism. Both techniques commonly fixate at less than 3 mm posterior to the limbus, within the pars plicata, increasing the risk of macular edema and vitreous hemorrhage.13,14
In practice, SSF is often performed using polypropylene or expanded polytetrafluoroethylene (Gore-Tex, W.L. Gore) sutures, with the latter gaining favor for its durability.15 IOLs with eyelets or closed-loop haptics allow for more controlled suture passage and fixation, although surgeons may also secure three-piece lenses by looping sutures around the haptics, when needed.
The main limitation tends to emerge later in the postoperative course. Suture erosion or breakage can lead to delayed decentration or dislocation, which tends to appear years after surgery.16
ISHF techniques, including the Yamane technique and its variants, rely on externalizing the haptics and tucking them into scleral tunnels (Figure).17 Compared with SSF, ISHF reduces conjunctival manipulation and eliminates suture-related complications, such as erosion and late breakage. IOL selection—typically three-piece monofocal IOLs—can influence intraoperative handling and postoperative stability. Lenses with polyvinylidene fluoride haptics, such as the CT Lucia 602 (Carl Zeiss Meditec), are often favored because of their flexibility and resistance to permanent deformation during externalization and flanging. Other common three-piece platforms include the Sensar AR40e and Tecnis ZA9003 (Johnson & Johnson Vision), LI61AO (Bausch + Lomb), and AcrySof MA60AC (Alcon).18-20 Dedicated scleral-fixation designs, such as the Carlevale IOL (FIL-SSF, Soleko), incorporate sutureless anchoring elements to improve stability and reduce tilt. The Carlevale IOL is not available in the United States, and although early series and a recent meta-analysis describe generally favorable outcomes, complication rates remain significant, in part due to its fixation in the pars plicata.21,22
Reports of postoperative optic–haptic junction instability have led some surgeons to reconsider haptic material and stiffness when selecting lenses for ISHF.23,24 Thus, some now select lenses with stiffer PMMA haptics for ISHF, despite the added technical difficulty during haptic externalization.
Figure. During the 27-gauge Yamane technique, the surgeon uses forceps to grab the IOL haptic (A), externalize it (B), and then use cautery to create a flange (C).
COMPLICATION PATTERNS ACROSS TECHNIQUES
Although complication rates are broadly similar when techniques are appropriately selected, the incidence remains frustratingly higher compared with other vitreoretinal procedures.25,26 Differences in complication profiles between techniques emerge in timing and underlying mechanism.27
ACIOL-related complications tend to appear later and are largely driven by chronic interaction with the corneal endothelium and angle structures.28,29
Iris-fixated lens approaches fall somewhere between mechanical and inflammatory domains, with pigment dispersion, low-grade inflammation, and dysphotopsias representing the most common concerns.30
Scleral-fixated lens techniques show a bimodal pattern. Early complications are usually related to surgical manipulation (eg, hemorrhage or malposition), whereas later issues depend on the fixation method itself: suture failure in SSF and geometric or alignment-related problems in ISHF.31-33
Across all approaches, complications tend to be mechanism-specific failure modes rather than uniform procedural risks. In addition, intraoperative challenges are significant as techniques rely on ad hoc modifications and off-label use of lenses, leading to breakage and distortion. Thus, these cases tend to have prolonged surgical time and have become less profitable for surgery centers and surgeons alike.34
Recent cost analyses indicate that secondary IOL surgeries—particularly those requiring combined vitrectomy and multiple surgeons—often generate negative day-of-surgery margins under Medicare reimbursement, with single surgeon vitrectomy cases covering approximately 87% to 92% of costs and multiple-surgeon vitrectomy cases covering approximately 56% to 63%.35 Despite these economic challenges, some surgeons have highlighted secondary IOL surgery as a potential practice-builder, emphasizing its role in expanding referral networks and providing comprehensive care for complex aphakic and subluxated-lens patients.36,37
AN INVESTIGATIONAL SECONDARY IOL APPROACH
A prosthetic capsular bag (LensOne, Long Bridge Medical), designed to accommodate a range of one- and three-piece IOLs, received FDA Breakthrough Device Designation. The company also received FDA clearance to initiate the Golden Gate pivotal trial, which will evaluate the safety and effectiveness of the prosthetic capsular bag in patients with IOL dislocation who are poor candidates for an anterior-chamber IOL.1
In a first-in-human study of 15 patients who underwent secondary IOL surgery with the prosthetic capsular bag, 14 were evaluable at 12 months: 71% had an uncorrected VA within 2 lines of the screening BCVA, and 93% had a BCVA within 1 line. The one patient lost to follow-up had met both endpoints at 3 months. While the 12-month safety profile was reassuring, longer-term follow-up is necessary, the researchers concluded in their paper.2
1. FDA Clears Long Bridge Medical to Initiate US Pivotal Clinical Study for LensOne Prosthetic Capsular Bag [press release]. Eyewire+ March 27, 2026. Accessed August 26, 2026. tinyurl.com/kxph68nm
2. Brodie F, Naseri A, Ho I, et al. Clinical outcomes of a prosthetic capsular bag for scleral fixation of an intraocular lens in eyes without capsular support. Ophthalmology. 2026;133:965-970. doi.org/10.1016/j.ophtha.2026.04.021
PRACTICAL APPROACH TO SECONDARY IOL SELECTION
Secondary IOL implantation is increasingly guided by individualized surgical planning rather than fixed procedural preference. While visual outcomes are broadly comparable across techniques, differences in complication mechanisms, recovery profiles, surgeon comfort, and technical demands remain central to decision making.
1. Dabrowska-Kloda K, Kloda T, Boudiaf S, Jakobsson G, Stenevi U. Incidence and risk factors of late in-the-bag intraocular lens dislocation: Evaluation of 140 eyes between 1992 and 2012. J Cataract Refract Surg. 201;41(7):1376. doi.org/10.1016/j.jcrs.2014.10.040
2. Mönestam E. Frequency of intraocular lens dislocation and pseudophacodonesis, 20 years after cataract surgery – a prospective study. Am J Ophthalmol. 2019;198:215-222. doi.org/10.1016/j.ajo.2018.10.020
3. Chen SP, Chang DF. Secondary IOL implantation without capsular support. Am J Ophthalmol. 2025;280:230-237. doi.org/10.1016/j.ajo.2025.08.023
4. Shen JF, Deng S, Hammersmith KM, et al. Intraocular lens implantation in the absence of zonular support: an outcomes and safety update: a report by the American Academy of Ophthalmology. Ophthalmology. 2020;127(9):1234-1258. doi.org/10.1016/j.ophtha.2020.03.005
5. Vounotrypidis E, Schuster I, Mackert MJ, Kook D, Priglinger S, Wolf A. Secondary intraocular lens implantation: a large retrospective analysis. Graefes Arch Clin Exp Ophthalmol. 2019;257(1):125-134. doi.org/10.1007/s00417-018-4178-3
6. Bantounou MA, Aggarwal N, Boucher N, Sharma C, Starr MR. Trends and outcomes of secondary intraocular lens utilization among vitreoretinal surgeons compared to vitreoretinal and anterior segment surgeons. Ophthalmic Surg Lasers Imaging Retina. 2025;56(10):602-607. doi.org/10.3928/23258160-20250725-01
7. Jonker SMR, Berendschot TTJM, Ronden AE, Saelens IEY, Bauer NJC, Nuijts RMMA. Five-year endothelial cell loss after implantation with Artiflex Myopia and Artiflex Toric Phakic intraocular lenses. Am J Ophthalmol. 2018;194:110-119. doi.org/10.1016/j.ajo.2018.07.015
8. Kent C. No capsular support: Do ACIOLs still make sense? Review of Ophthalmology. 2018;25(5):52-58
9. Saad EME, Mohammed HEM, Tharwat E, et al. Visual outcome, safety, and complications of iris suture fixation versus iris claw fixation for aphakic correction: a prospective randomized study. BMC Ophthalmol. 2026;26:177. doi.org/10.1186/s12886-026-04785-x
10. Choi EY, Lee CH, Kang HG, et al. Long-term surgical outcomes of primary retropupillary iris claw intraocular lens implantation for the treatment of intraocular lens dislocation. Sci Rep. 2021;11:726. doi.org/10.1038/s41598-020-80292-3
11. Schallenberg M, Dekowski D, Hahn A, Laube T, Steuhl KP, Meller D. Aphakia correction with retropupillary fixated iris-claw lens (Artisan) – long-term results. Clin Ophthalmol. 2014;8:137-141. doi.org/10.2147/OPTH.S55205
12. Cakir I, Demir G, Yildiz BK, Öcal MC, Yildirim Y, Agca A. Efficacy and safety of iris-supported phakic lenses (Verisyse) for the treatment of high myopia: 5-year results. Int Ophthalmol. 2021;41(8):2837-28-45. doi.org/10.1007/s10792-021-01841-x
13. Hauser D, Ben-David D, Masarwa D, et al. Cystoid macular edema after four-point scleral fixation of intraocular lens. Retina. 2021;41(10):2035-2040. doi.org/10.1097/IAE.0000000000003142
14. Day HR, Durrani AK, Kim SJ, Patel S. Outcomes and complications of concurrent pars plana vitrectomy and scleral-fixated intraocular lens placement using Gore-Tex suture. J Vitreoretin Dis. 2019;4(2):119-124. doi.org/10.1177/2474126419895691
15. Khan MA, Gupta OP, Smith RG, et al. Scleral fixation of intraocular lenses using Gore-Tex suture: clinical outcomes and safety profile. Br J Ophthalmol. 2016;100(5):638-643. doi.org/10.1136/bjophthalmol-2015-306839
16. Kokame GT, Yanagihara RT, Shantha JG, Kaneko KN. Long-term outcome of pars plana vitrectomy and sutured scleral-fixated posterior chamber intraocular lens implantation or repositioning. Am J Ophthalmol. 2018;189:10-6. doi.org/10.1016/j.ajo.2018.01.034
17. Yamane S, Sato S, Maruyama-Inoue M, Kadonosono K. Flanged intrascleral intraocular lens fixation with double-needle technique. Ophthalmology. 2017;124(8):1136-1142. doi.org/10.1016/j.ophtha.2017.03.036
18. Czajka MP, Frajdenberg A, Stopa M, Pabin T, Johansson B, Jakobsson G. Sutureless intrascleral fixation using different three-piece posterior chamber intraocular lenses: a literature review of surgical techniques in cases of insufficient capsular support and a retrospective multicentre study. Acta Ophthalmologica. 2020;98(3):224-236. doi.org/10.1111/aos.14307
19. Ma KK, Yuan A, Sharifi S, Pineda R. A biomechanical study of flanged intrascleral haptic fixation of three-piece intraocular lenses. Am J Ophthalmol. 2021;227:45-52. doi.org/10.1016/j.ajo.2021.02.021
20. Icoz M, Erkan E. Comparison of two different intraocular lenses used in the modified Yamane technique [published online ahead of print June 4, 2025]. J Cataract Refract Surg. doi.org/10.1097/j.jcrs.0000000000001701
21. Carlà MM, Boselli F, Giannuzzi F, et al. Sutureless scleral fixation Carlevale IOL: a review on the novel designed lens. Int Ophthalmol. 2023;43(6):2129-2138. doi.org/10.1007/s10792-022-02579-w
22. Ripa M, Angunawela R, Motta L. Scleral fixation of carlevale intraocular lens: A systematic review and meta-analysis. Retina. 2023;43(10):1750. doi.org/10.1097/IAE.0000000000003873
23. Safran JP, Safran SG. Intraocular lens tilt due to optic-haptic junction distortion following intrascleral haptic fixation with the Yamane technique. Am J Ophthalmol Case Rep. 2023;30:101845. doi.org/10.1016/j.ajoc.2023.101845
24. Icoz M, Erkan E, Kılınc AK. Comparison of polyvinylidene fluoride and poly(methyl methacrylate) haptics used in the modified Yamane technique. J Cataract Refract Surg. 2026 Aug 1;52(8):737-743. doi.org/10.1097/j.jcrs.0000000000001914
25. Kaufmann GT, Gupta O, Yu J, et al. Vitreoretinal outcomes following secondary intraocular lens implantation with pars plana vitrectomy. Retina. 2024;44(8):1337-1343. doi.org/10.1097/IAE.0000000000004139
26. Kaufmann GT, Boucher N, Sharma C, Starr MR. Trends in secondary intraocular lens surgery among vitreoretinal surgeons. Ophthalmol Retina. 2023;7(11):965-971. doi.org/10.1016/j.oret.2023.07.008
27. Stem MS, Todorich B, Woodward MA, Hsu J, Wolfe JD. Scleral-fixated intraocular lenses: past and present. J Vitreoretin Dis. 2017;1(2):144-1452. doi.org/10.1177/2474126417690650
28. Huang Y sen, Xie L xin, Wu X ming, Han D sheng. [Long-term follow-up of flexible open-loop anterior chamber intraocular lenses implantation]. Zhonghua Yan Ke Za Zhi. 2006;42(5):391-395.
29. Ravalico G, Botteri E, Baccara F. Long-term endothelial changes after implantation of anterior chamber intraocular lenses in cataract surgery. J Cataract Refract Surg. 2003;29(10):1918-1923. doi.org/10.1016/s0886-3350(02)02052-7
30. Gonnermann J, Klamann MKJ, Maier AK, et al. Visual outcome and complications after posterior iris-claw aphakic intraocular lens implantation. J Cataract Refract Surg. 2012;38(12):2139-2143. doi.org/10.1016/j.jcrs.2012.07.035
31. Patel NA, Shah P, Yannuzzi NA, et al. Clinical outcomes of 4-point scleral fixated 1-piece hydrophobic acrylic equiconvex intraocular lens using polytetrafluoroethylene suture. Clin Ophthalmol. 2018;12:2145-2148. doi.org/10.2147/OPTH.S174211
32. Medin H, Kure ISH, Dalby M, Drolsum L, Kristianslund O. Long-term risk of suture breakage after scleral fixation of late in-the-bag intraocular lens dislocation with a polypropylene 10-0 suture. J Cataract Refract Surg. 2022;48(12):1408-1412. doi.org/10.1097/j.jcrs.0000000000001010
33. Lin H, Ye X, Huang X, et al. Long-term stability of intraocular lens with trimmed or untrimmed haptics in Yamane sutureless intrascleral fixation technique. Med Sci Monit. 2021;27:e928868. doi.org/10.12659/MSM.928868
34. Meshkin RS, Hoffman SE, Chaaya MD C, et al. Time-driven, activity-based cost analysis of secondary intraocular lens implantation. J Vitreoretin Dis. 2026;24741264261428752. doi.org/10.1177/24741264261428752
35. Pan WW, Young C, Portney D, et al. Comparative cost analysis of secondary intraocular lens surgeries using time-driven activity-based costing. Am J Ophthalmol. 2025;273:159-66. doi.org/10.1016/j.ajo.2025.02.008
36. Stevenson S. ASRS 2023: Tracking trends in secondary IOL surgery among vitreoretinal surgeons. Ophthalmology Times. July 31, 2023. Accessed September 14, 2026. tinyurl.com/46xrkyhd
37. Adam M. Synergizing anterior and posterior expertise. Cataract & Refractive Surgery Today. March 14, 2024. Accessed September 14, 2026. crstoday.com/crst-issues/mar-2024/synergizing-anterior-and-posterior-expertise/45242